Preparing For Use
This chapter gives you instructions for preparing the logic analyzer for use.
Power Requirements
The logic analyzer requires a power source of either 115 Vac or 230 Vac,
Operating Environment
The operating environment is listed in chapter 1. Note the noncondensing humidity limitation. Condensation within the instrument can cause poor operation or malfunction. Provide protection against internal condensation.
The logic analyzer will operate at all specifications within the temperature and humidity range given in chapter 1. However, reliability is enhanced when operating the logic analyzer within the following ranges:
•Temperature: +20 °C to +35 °C (+68 °F to +95 °F)
•Humidity: 20% to 80% noncondensing
Storage
Store or ship the logic analyzer in environments within the following limits:
•Temperature:
•Humidity: Up to 90% at 65 °C
•Altitude: Up to 15,300 meters (50,000 feet)
Protect the logic analyzer from temperature extremes which cause condensation on the instrument.
To inspect the logic analyzer
1Inspect the shipping container for damage.
If the shipping container or cushioning material is damaged, keep them until you have checked the contents of the shipment and checked the instrument mechanically and electrically.
2Check the supplied accessories.
Accessories supplied with the logic analyzer are listed in "Accessories" in chapter 1.
3Inspect the product for physical damage.
Check the logic analyzer and the supplied accessories for obvious physical or mechanical defects. If you find any defects, contact your nearest Agilent Technologies Sales Office. Arrangements for repair or replacement are made, at Agilent Technologies’ option, without waiting for a claim settlement.