Testing Performance

To test the single-clock, multiple-edge, state acquisition

2Check the data pulse width. Using the oscilloscope, verify that the data pulse width is 4.000 ns, +0 ps or 100 ps.

a In the oscilloscope Timebase menu, select Sweep Speed 1.00 ns/div.

b Select Delay. Using the oscilloscope knob, position the data waveform (Channel 1) so that the waveform is centered on the screen.

cIn the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set Marker 1 at 1.3000 V. Set the Marker 2 Position to Chan 1, then set Marker 2 at 1.3000 V.

dIn the oscilloscope Delta T menu, select Start On Pos Edge 1. Select Stop on Neg Edge 1. Select Precision Edge Find.

eIf the pulse width is outside of the limits, adjust the pulse generator channel 1 width and select the oscilloscope Precision Edge Find until the pulse width is within limits.

3–50