Troubleshooting

To test the logic analyzer probe cables

4Set up the Format menu. a Press the Format key.

b Move the cursor to the field showing the channel assignments for the pod under test. Press the Clear Entry key until the pod channels are all assigned (all asterisks (*)). Press the Done key.

cSelect Master Clock, then select a double edge for the clock of the pod under test. Turn off the other clocks.

dIn the Master Clock menu, select Setup/Hold, then select 4.0/0.0 ns for the pod being tested. Select Done. Select Done again to exit the Master Clock menu.

5–29

Page 117
Image 117
Agilent Technologies 01664-97005 manual Set up the Format menu. a Press the Format key