2Set up the Format menu.
aPress the Format key. Select State Acquisition Mode, then select Full Channel/4K
Memory/50MHz.
bSelect the field to the right of each Pod field, then select ECL.
3Set up the Trigger menu.
aPress the Trigger key. Select Clear Trigger, then select All.
bSelect the Count Off field, then select Time in the pop-up menu. Select Done to exit
the menu.
cSelect the field labeled 1 under the State Sequence Levels. Select the field labeled
"anystate", then select "no state." Select Done to exit the State Sequence Levels menu.
dSelect the field next to the pattern recognizer "a," under the label Lab1. Type "000A" ,
then press Select.
Testing Performance
To test the multiple-clock, multiple-edge, state acquisition
3–36