cAdjust the pulse generator channel 1 Delay, then select Precision Edge Find in the
oscilloscope Delta T menu. Repeat this step until the pulses are aligned according to
the setup time of the setup/hold combination selected, +0.0 ps or 100 ps.
4Select the clock to be tested.
aIn the Master Clock menu, select the clock field to be tested, then select the clock
edge as indicated in the table. The first time through this test, use the top clock and
edge in the following table.
Clocks
J
K
bConnect the clock to be tested to the pulse generator channel 2 output.
cSelect Done to exit the Master Clock menu.
Testing Performance
To test the single-clock, single-edge, state acquisition
3–30