ACPGR Adjacent Channel Power Graph On or Off
Setting the ACPGR to
-
1, does the following:
n Turns off the display of the ACP graph in trace C.
n
Calculates and displays the ACP numeric results.
n Removes the
GRPH
MKR
ON OFF softkey from the menu.
n
Turns off the ACP graph marker.
n If in continuous measurement mode, the numeric results will be updated at the end of every
sweep.
When the graph is generated, the reference level represents an adjacent channel power ratio of
0, and the horizontal center represents a frequency offset (and channel spacing) of 0 Hz. The
results of ACPGR are not defined for frequency spacings where the graph is drawn below the
bottom graticule line.
Programming Commands 5-41