MKA Marker Amplitude
Example
OUTPUT
718;"TDF
P;AUNITS
DBM;MKA?;"
This returns the amplitude value of the marker
(amplitude unit is
dBm).
If the trace data format is used with trace data format A, the result depends on the setting of
the MDS command.
Example
OUTPUT
718;"TDF
A;MDS
B;MKA?;"
Returns one byte representing the marker vertical
position (-32,768 to 32,767) divided by 32 and then
AhDed
with 255.
OUTPUT
718;"TDF
A;MDS
H;MKA?;"
Returns two bytes in a binary word format that has
a value from -32,768 to 32,767. The value represents
the binary trace amplitude value.
Using the trace data format I is equivalent to the TDF A format. If the trace data format is
used with trace data format M, the result is returned in ASCII measurement units (-32,768 to
32,767).
Example
OUTPUT
718;"TDF
M;MKA?;"
Returns one vertical position value in measurement units.
Query Response
5-326 Programming Commands