FS Full Span

FS

Full Span

Sets the frequency span of the spectrum analyzer to full span.
Syntax
Equivalent Softkey: FULL SPAN .
Related Commands: CF, FA, FB, HNLOCK, SP, SS.
Example
OUTPUT 7 18 ; “FS
;
Puts the spectrum analyzer in full-span
mode.
Description
The FS command selects both the start frequency and the stop frequency according to the
frequency span of the spectrum analyzer. Resolution bandwidth, video bandwidth, and sweep
time are all set to autocoupled.
Spectrum Analyzer start
stop
Model
Frequency Frequency
HP 8590L or HP 85913
0 Hz
1.8 GHz
HP
8592L*
or HP
8593E* 2.750 GHz 22 GHz
HP 85943
0 Hz
2.9 GHz
HP
8595E*
0 Hz
6.5
GHz
HP 85963'
0
12.8
GHz
I
*In harmonic unlock.
I
In harmonic lock, “FS;” sets both the center frequency and the span according to the harmonic
band shown in the following table.
Center Frequency and Span Setting
2
HP
8592L,
HP 85933, or HP 85963 only.
Programming Commands