IP Instrument Preset
IP also clears all user graphics, all on-event algorithms, and turns off
the
windows display
mode. (The on-event algorithms are
ONCYCLE,
ONDELAY,
ONEOS, ONMKR, ONSRQ, ONSWP,
and
ONTIME.)
Instrument preset automatically occurs when you turn on the spectrum analyzer. IP is a good
starting point for many measurement processes. When IP is executed remotely, the spectrum
analyzer does not necessarily execute a complete sweep, however. You should execute a take
sweep (TS) to ensure that the trace data is valid after an IP.
Note that ONPWRUP will be terminated,
not
erased, by performing one of the following:
n
Pressing
[?Z5Z?-j
on
the spectrum analyzer.
n
Executing the universal HP-IB command DCL Device Clear.
n
Executing the RS-232 command BREAK.
To remove ONPWRUP execute the DISPOSE ONPWRUP command.
Programming Commands 5-245