LIMIFT

Select Frequency or Time Limit Line

Selects how the limit-line segments are defined: according to frequency, or according to the
sweep time setting of the spectrum analyzer.
Syntax
Equivalent
Softkey:
LIMIFT is equivalent to LIMITS
FRIJ
TIME.
Related Commands: LIMIDEL, LIMILINE, LIMIMODE, LIMIREL, LIMISEG, LIMISEGT,
SEGDEL, SENTER, SENTERT.
Example
OUTPUT
718;"LIMIFT
TIME;"
Ifthe
current
limit-line
table
is
a frequency limit-line table,
it is purged.
LIMIFT
TIME places the limit-line segments
on the spectrum analyzer display with respect to the sweep
time of the spectrum analyzer:
Description
If you execute “LIMIFT TIME;“, LIMISEGT, or SENTERT, the limit-line segments are placed on
the spectrum analyzer display with respect to the sweep time setting of the spectrum analyzer.
If you execute “LIMIFT FREQ;“, LIMISEG, or SENTER, the limit-line segments are placed
according to the frequency that is specified for each segment. If a limit line has already been
defined, changing the
LIMIFI’
setting clears the existing limit line.
Query Response
TIME
FREQ
5-274