LIMIFAIL

Limits Failed

Returns a “0” ifthelastmeasurementsweep of trace
AisequaIto
or within the limit-line
bounds.
Syntax
Pre’def
ined
function path only
Related Commands: LIMIHI, LIMILINE, LIMILO, LIMISEG, LIMISEGT, LIMITEST, SENTER,
SENTERT.
Example
10
OUTPUT
718;“IP;SNGLS;CF300MHZ;SPlOOMHZ;”
20
OUTPUT
718;“LIMIDEL;”
30
OUTPUT
718;“LIMIMODE
UPPER;”
40
OUTPUT
718;“LIMISEG
250MHZ,-60DB,FLAT;”
Initializes spectrum analyzer and
changes the frequency and span
settings.
Deletes any limit-line tables, sets
the table type to
$xed.
Spec@es
the upper limit-line table.
Creates an entry for the upper
limit-line table. Because the LIMISEG
command is used, the limit-line
will be boxed on the frequency.
OUTPUT 718;“LIMISEG
290MHZ,-60DB,SLOPE;”
OUTPUT
718;"LIMISEG
295MHZ,-lSDB,FLAT;"
OUTPUT
718;"LIMISEG
305MHZ,-15DB,SLOPE;"
OUTPUT
718;"LIMISEG
3lOMHZ,-GODB,FLAT;"
OUTPUT
718;"LIMISEG
SlOMHZ,-GODB,FLAT;"
OUTPUT
718;"LIMITEST
0N;TS;"
Turns on limit-line testing.
OUTPUT
718;"LIMIFAIL?;"
Returns the status of
the
limit-line
testing.
ENTER 718;A
DISP A
Displays the result.
END
50
60
70
80
90
100
110
120
130
140
5.272
Programming Commands