MKPK Marker Peak

MKPK

Marker Peak

Positions the active marker on a signal peak.
Syntax
MKPK
Equivalent
Softkey:
NEXT
PEAK , NEXT PK RIGHT,
I?EXT
PK LEFT, and [PEAK SEARCH].
Related Commands: MKCF, MKF, MKOFF, MKPX, TH.
Example
10
OUTPUT
718;"IP;"
20 OUTPUT 718;"SNGLS;TS;MKPK HI;"
30
OUTPUT
718;"MKA?;"
40
ENTER 718;A
50
DISP A
60
END
Description
Initializes the spectrum analyzer:
Places active marker on highest peak.
Returns
amplitude
value
of
marker
to the computer
Puts the spectrum analyzer response in the com-
puter variable, A.
tiplays
amplitude value.
Executing MKPK HI, or simply MKPK, positions the active marker at the highest signal
detected. If an active marker is on the screen, the MKPK parameters move the marker as
follows:
HI (highest)
moves the active marker to the highest peak.
NH
(next highest)
moves the active marker to the next signal peak of lower amplitude.
NR
(next right)
moves the active marker to the next signal peak of higher frequency.
NL (next left)
moves the active marker to the next signal peak of lower frequency.
Note
This function is for use with the frequency markers only.
Programming Commands
5-351