DN Down

DN

Down

Reduces the active function by the applicable step size.
Syntax
(
DN
XDN
Related Commands: See the list of active functions listed in the description for DN.
Example
OUTPUT 718;“SS
1MHZ;CF
1GHZ;DN;”
Setscenterfrequency to 1
GHz.
OUTPUT
718;“SP
40MHZ;MKPK;DN;”
Decreases the frequency span.
Description
Before executing DN, be sure that the function to be decreased is the active function. For
example, the second line of the programming example decreases the span, because marker
peak (MKPK) is not an active function.
The active functions are ACPBW, ACPSP, AT, CF, CRTHPOS, CRTVPOS, DL,
DOTDENS,
FA,
FB, FMGAIN, GD, GL, LG, MKA, MKD, MKFCR, MKN, MKPAUSE, MKPX, ML, NDB, NRL, RB,
RCLS, ROFFSET, RL, RLPOS, SAVES, SAVRCLN,
SETDATE,
SETTIME,
SP, SQLCH, SRCALC,
SRCAT, SRCPOFS, SRCPSWP, SRCPWR, SRCTK, SS, ST, TH, TVLINE, VB, VBR, and user-defined
active function specified by the ACTDEF command.
Programming Commands
5-137