MEASOFF’ Measurement Off

MEASOFF

Measurement Off

Turns off the current measurement, erases the display, and then displays the
MEAS/USER
menu.
Syntax
Equivalent
Softkey:
MEBS
OFF
.
Related Commands: ACP, ACPE, ACPGRAPH, CHP, OBW.
Example
OUTPUT
718;"MEASOFF;"
lbns
ofl the current measurement.
Description
If ACPPAR is set to automatic, executing MEASOFF returns the following spectrum analyzer
settings back to their premeasurement settings:
H
Frequency span, resolution bandwidth, video bandwidth, center frequency step size, and
sweep time.
n
Detector mode.
n
Amplitude scale.
MEASOFF does
not
do any of the following:
n
Change the values of
the
channel spacing (ACPSP) or the channel bandwidth (ACPBW).
n Restore the trace contents, trigger mode, amplitude units, and any trace math functions (see
‘l%ble
5-4 for a list of the trace math functions) to their premeasurement state.
You should execute MEASOFF when you no longer want to use any of the power menu
functions. (The power menu functions are ACP, ACPE, ACPGRAPH, CHP, and OBW).
MEASOFF may also turn off some other spectrum analyzer measurements. For example,
MEASOFF also turns off FFTCONTS, FFTAUTO, FFTSNGLS, NDBPNT, PCTAM, GDRVUTIL, and
TOI.
Programming Commands
5-309