RCLT

Recall Trace

Recalls previously saved trace data, amplitude factors, or limit-line data from the specified
trace register in spectrum analyzer memory. Trace data is recalled with instrument state, date,
and screen title.
Syntax
RCLT
user-defined trace
trace range
I
XRCLl
Item Description/Default Range
User-defined trace
A
trace defined by the TRDEF command. Any valid trace name.
Trace range
A segment of trace A, trace
B,
trace
C,
or a user-defined trace.
Number Any valid integer. 0 to TRCMEM
-
1.
Equivalent Softkey: The RCLT command and the VIEW commands are equivalent to
Intsmal
-2
Trace.
Parameter Value: 0 to TRCMEM
-
1.
Prerequisite Commands: TRDEF when using a user-defined trace.
Related Commands: CAT, CLRW, LOAD, SAVET, SNGLS, TRCMEM, TS, VIEW.
Example
10 OUTPUT
718;"IP;CF
300MHZ;SP
20MHZ;TS;"
Initializes spectrum
anulyq
changes
the center frequency and span.
20 OUTPUT
718;"SAVET
TRA,l;"
30 OUTPUT
718;"IP;"
40 OUTPUT
718;"RCLT
TRA,l;VIEW
TRA;"
50
END
Saves spectrum analyzer state and trace
A
data
in register 1.
Initializes spectrum analyzer
Recalls
spectrum analyzer state, trace
data; displays the result.
5-460
Programming Commands