GDRVCLPAR

Clear Pulse Parameters

Clears the pulse parameters (pulse width, pulse repetition interval, and reference edge) for a
time-gate measurement by setting the pulse parameters to 0.
Syntax
XGDRVCLPAR
Equivalent
Softkey:
CLE&R
PARAM
.
Option Required: Option 105. Option 101 is recommended.
Related Commands: GDRVPWID, GDRVPRI, GDRVREFE, GDRVST, GDRVVBW, GDRVRBW.
Example
OUTPUT
718;"GDRVCLPAR;"
Clears all the pulse parameters.
Description
GDRVPWID, GDRVPRI, GDRVREFE are the programming commands that can be used to set the
pulse width, pulse repetition interval, and reference edge, respectively.
GDRVCLPAR also turns off the resolution bandwidth to pulse width coupling, video bandwidth
to gate length coupling, and sweeptime to pulse repetition interval coupling.
The GDRVCLPAR command is a gate utility function (see “GDRVUTIL” for more information
about the gate utility functions). Executing a gate utility command without Option 105
installed in the spectrum analyzer generates the
COMMAND ERROR:
message and an SRQ 140.
5-180
Programming Commands