LIMITEST Enable Limit Line Testing

LIMITEST

Enable Limit Line ‘&sting

Compares trace A with the current limit-line data.
Syntax
XLIMITEST
Preset State: LIMITEST OFF.
Related Commands: LIMIFAIL, LIMIHI, LIMILO, LIMISEG, LIMISEGT, SENTER,
SENTERT.
Restrictions: Not available with Analog+ display mode. See “ANLGPLUS” for more
information.
Example
10
OUTPUT
7l8;"IP;SNGLS;CF300MHZ;SPlOOMHZ;"
20
OUTPUT
718;"LIMIDEL;"
30
OUTPUT
718;"LIMIMODE
UPPER;"
35
OUTPUT
718;"LIMIFT
FREQ;"
40
OUTPUT
718;"LIMISEG
250MHZ,-60DB,FLAT;"
50
OUTPUT
718;"LIMISEG
290MHZ,-60DB,SLOPE;"
60
OUTPUT
718;"LIMISEG
295MHZ,-lSDB,FLAT;"
70
OUTPUT
718;"LIMISEG
305MHZ,-lSDB,SLOPE;"
80
OUTPUT
718;"LIMISEG
310MHZ,-GODB,FLAT;"
90
OUTPUT
718;"LIMISEG
SlOMHZ,-GODB,FLAT;"
100
OUTPUT
718;"LIMITEST
0N;TS;"
110
OUTPUT
718;"LIMIFAIL?;"
120
ENTER
718;A
130
DISP A
140
END
Initializes spectrum analyzer and
changes the frequency and span
settings.
Deletes any limit-line tables,
sets
the table type to
jixed.
Specifies the upper limit-line table.
Selects a limit line
based
onfrequency.
Creates an entry to the upper
limit-
line table.
Turns on limit-line testing.
Returns the status of the limit-line
testing.
Displays
the result.
Programming Commands