ONPWRUP On Power Up

ONPWRUP

On Power Up

Executes the list of spectrum analyzer commands once on power up.
Syntax
/
commands to
/
be executed
\
Related Commands: CAT, DISPOSE, ERASE,
ONDELAY,
ONEOS, ONMKR, ONSWP,
ONTIME
Example
10 OUTPUT
718;"0NPWRUP!
SPlOMZ;
CF300MZ;!;"
Sets the spectrum
analyzer
span to
10
MH..
and the center frequency to
3OOMHizuponinstrumentpowwup.
Description
Limit the number of characters (between the delimiters) in the list of spectrum analyzer
commands to a maximum of 2047 characters.
Clearing the ONPWRUP definition: DISPOSE.
Note
If an infinite loop is inadvertently programmed in an ONPWRUP, press [PRESET)
to terminate the ONPWRUP, then DISPOSE the ONPWRUP. Powering the
spectrum analyzer on and off will not terminate the ONPWRUP.
Programming Commands
5-401