NRL

Normalized Reference Level

Sets the normalized reference level.
Syntax
Item
Description/Default
I
Range
Number Any real or integer number. Default unit is dBm.
Preset State: 0
dB.
Related Commands: MEASURE, RL, RLPOS.
Example
10
OUTPUT
718*"IP*"
20
OUTPUT
718;"SRCPWR
-1ODB;"
30
PRINT "CONNECT TRACKING GENERATOR OUTPUT TO INPUT"
40 PRINT "THEN PRESS CONTINUE"
50
PAUSE
60
OUTPUT
718;"MEASURE
SR;"
70
OUTPUT
718;"CLRW
TRB;TS;"
80
OUTPUT
718;"BLANK
TRB;"
so
PRINT "CONNECT DEVICE TO ANALYZER"
100
PRINT "THEN PRESS CONTINUE"
110
PAUSE
120
OUTPUT
718;"CLRW
TRA;TS;"
130
OUTPUT
718;"AMBPL
ON;"
140
OUTPUT
718;"RLPOS
4;"
150
OUTPUT 718;"NRL
40DB;"
160
END
Initializes spectrum
anal2/2er
Sets tracking generator output.
Sets the measurement mode
to stimulus-response.
lbkes
a measurement sweep
of thejlatness of the system.
Zxkes
a mxxzsurement sweep
of the device under test in
the system.
Normalizes the response.
Changes the location of the
reference level to thefourth
graticule
from the bottom.
Oflsets
trace A so it is
on-
screen.
5-384 Programming Commands