SRCPSTP Source Power-Level Step Size

SRCPSTP

Source Power-Level Step Size

Selects the source-power step size.
Syntax
I-
step size
XSRCPSTP
Item Description/Default Range
Number Any real or integer number.
Option Required: Option 010 or 011.
Equivalent
Softkey:
SRC
PWR STP SIZE.
Step Increment: 0.1
dB.
Related Commands: SRCPWR, SRCPOFS, SRCPSWP
Preset State: SRCPSTP AUTO (one major vertical ‘scale division).
Example
Select incremental changes of power effected by “SRCPWR UP;“, “SRCPWR DN;” commands, or
the step keys.
OUTPUT
718;"MEASURE
SR;"
Activates stimulus-response mode.
OUTPUT
718;"SRCPWR
-1ODB;"
Turns on the source output.
OUTPUT
718;"SRCPSTP
.3DB;"
Sets power-level step
size
to 0.3
dB
OUTPUT
718;"SRCPWR
UP;"
Increases the power level.
Description
The SRCPSTP command selects the step size for the following source commands:
n
Power offset (SRCPOFS).
n
Power sweep (SRCPSWP).
n
Power (SRCPWR).
Use SRCPSTP to set the step size to a specific value.
“SRCPSTP AUTO;” sets the step size to one vertical scale division.
Programming Commands 5-511