ACPPAR

ACP Manual or Auto

Determines if the spectrum analyzer settings used for the adjacent channel power (ACP),
adjacent channel power extended (ACPE), channel power (CHP), or occupied bandwidth (OBW)
measurement are set manually or automatically.
Syntax
Equivalent Softkey:
PARA?!
AUTO MAN .
Related Commands: ACP, ACPE, ACPSP, ACPBW, OBW, CHP, DET, ST, VB, RB, SS.
Example
OUTPUT
718;"MOV
ACPPAR,l;"
Sets theparametersfor the adjacent channel
powerparam-
eters
automatically.
Description
When ACPPAR is set to 1, the spectrum analyzer settings for the ACP, ACPE, CHP, or OBW
measurement are set by the spectrum analyzer. When ACPPAR is set to 1, the spectrum
analyzer does the following before the measurement is performed:
n
Performs the trace preset (TRPRST) command.
n Changes the trigger mode to free run.
n
Changes the detector mode to sample.
n
Changes the amplitude scale to 10
dB
per division.
n Sets the frequency span, resolution bandwidth, video bandwidth, center frequency step size,
and sweep time based on the channel spacing (ACPSP) and channel bandwidth (ACPBW).
See the following table for more information about the spectrum analyzer settings for each
measurement.
n
‘fakes
a sweep.
5-44