ONSWP On Sweep
Description
The list of analyzer commands should not include a take sweep (TS). Limit the number of
characters (between the delimiters) in the list of spectrum analyzer commands to a maximum
of 2047 characters.
The #A, msb length, lsb length, and character data form an A-block data field. The A-block
data field is used when the length of the character data is known. To use the A-block data
format, #A must precede the msb length and
lsb
length. The msb length and lsb length
represent the length of the character data. The #I, character data, and EOI (END) form an
I-block data field. The I-block data field is used when the length of the character data is
unknown. The I-block data field is available for the HP-IB interface only.
Clearing the ONSWP definition: IP clears the ONSWP definition. You can use the DISPOSE
command to clear the ONSWP definition also.
Query Response

+(

#A

msb length
Isb
length analyzer
comnand
QONSWP
Programming Commands
5-405