SRCPOFS

Source Power Offset

Offsets the source power level readout.
Syntax
XSRCPOFS
Item Description/Default Range
Number Any real or integer number.
Option Required: Option 010 or 011.
Equivalent
Softkey:
$QlC
PWR OFFSET .
Related Commands: SRCPWR,
SRCPSWP
Step Increment: Determined by SRCPSTP
Preset State: 0
dB.
Example
Use SRCPOFS to offset the power-level readout for the tracking-generator source.
OUTPUT
718;"MEASURE
SR;"
Sets spectrum analyzer to stimulus-response
mode.
OUTPUT
718;"SRCPWR
-1ODB;"
Turns on source output.
OUTPUT
718;"SRCPOFS
13DB;"
Oflets
power-level
readout
for source by 13
dR
Description
The SRCPOFS command offsets the displayed power of the built-in tracking generator. This
function may be used to take into account system losses (for example, cable loss) or gains (for
example, preamplifier gain) reflecting the actual power delivered to the device under test.
Query Response
QQ,
5.510
Programming Commands