SRCPSWP Source Power Sweep

SRCPSWP

Source Power Sweep

Selects the sweep range of the source output,
Syntax
(
SRCPSWP
>
/-
sweep
--\
DB
I
I
I
I
I
.
.
.
.
.
.
.
.
.
.
.
.
.._._.............
.’
J
XSRCPSWP
Item Description/Default
R-w
Number Any real or integer number.
Option Required: Option 010 or 011.
Equivalent
Softkey:
PWR
SWP
ON OFF .
Step Increment: Determined by SRCPSTP.
Related Commands: SRCPSWP, SRCPOFS, SRCPSTP
Preset State: SRCPSWP OFF.
Example
Use SRCPSWP to sweep the power level of the source output.
OUTPUT
718;"MEASURE
SR;"
Activates stimulus-response mode.
OUTPUT
718;"SRCPWR
-1ODB;"
Sets power level of source output to -10
dBm.
OUTPUT
718;"SP
0;"
Sets span to 0 Hz.
OUTPUT 718;"SRCPSWP
IODB;"
Sweeps source output from -10 dBm to 0
dBm.
Description
The SRCPSWP command works in conjunction with the SRCPWR (source power) command to
sweep the amplitude level of the source output. The SRCPWR setting determines the amplitude
level at the beginning of the sweep. The SRCPSWP command determines the change in
amplitude level of the sweep.
For example, if SRCPWR and SRCPSWP are set to -15
dBm
and 4
dB
respectively, the source
sweeps from
-
15
dBm
to
-
11
dBm.
Programming Commands