RL

Reference Level

Specifies the amplitude value of the reference level.
Syntax
RL
-
ompl
itude
value
7
XRL
Item Description/Default Range
Number Any real or integer number. Default unit is the current
Amplitude range of
amplitude unit. the spectrum analyzer.
Equivalent
Softkey:
REF
LYL
.
Preset State: 0
dBm.
Step Increment: by 10
dBm.
Related Commands: AT, MKRL, ML, RESETRL, RLPOS.
Example
10 OUTPUT
718;"IP;SNGLS;CF
300MHZ;SP
20MHZ;"
20 OUTPUT
718;"TS;MKPK
HI;MKRL;TS;"
30 OUTPUT
718;"RL?;"
40 ENTER
718;RefJevel
50 PRINT "REFERENCE LEVEL
IS",Ref-level,"DM"
60 END
Initializes spectrum
analy,wq
acti-
vates single-sweep
mooTq
changes
cen-
tecfmww4,
span.
Zxkes
sweep, places marker on signal
peak, sets reference level to marker
level.
Queries reference level.
Puts the spectrum analyzer response
in the computer variable,
Reflevel.
5468
Programming Commands