INZ Input Impedance

INZ

Input Impedance

Specifies the value of input impedance expected at the active input port.
Syntax
Equivalent
Softkey:
fIJKIl’
Z
508
75Q
.
Preset Value: Returns impedance to the power-on value.
Related Commands:
AUNITS.
Example
OUTPUT
718;"INZ
75;"
Changes input impedance to 75 ohms.
OUTPUT 718
;
"AUNITS V;
Changes amplitude units to volts.
Description
The actual impedance can be affected only by internal hardware. With the exception of Option
001 or 011 (HP 859OL or HP 85913 only), the spectrum analyzer hardware supports
5Ofl
only.
The INZ command is used for computation purposes during power or voltage conversions.
The INZ command affects only the amplitude results that are reported in absolute relative
power
(dBm
units or watts). (See “AUNITS.“)
Query Response
QINZ
Programming Commands 5-241