SRCTK Source Tracking

SRCTK

Source Tracking

Adjusts the tracking of the source output with the spectrum analyzer sweep.
Syntax
SRCTK
Item
Description/Default
tinge
Number Any real or integer number.
0 to 16,383.
Option Required: Option 010 or 011.
Equivalent
Softkey:
Ml?
TM
ADJUST .
Step Increment: 1.
Related Commands: Commands that change bandwidth, such as RB, VB, ST, and commands
that change frequency, such as SP, CF, FA, FB, SP, FS, SRCTKPK.
Example
OUTPUT
718;"MEASURE
SR;"
Activates the stimulus-response
moo%?.
OUTPUT
718;"SRCPWR
-2ODB;"
Turns on the power at the source output.
OUTPUT
718;"SP
IMHZ;"
Sets measurement range.
OUTPUT
718;"RB
IKHZ;"
OUTPUT
718;"TS;"
Zzkes sweep.
OUTPUT
718;"SRCTK
EP;"
Allows entry
offrom
front-panel keys to adjust tracking.
Description
The SRCTK command adjusts the tracking of the tracking-generator output relative to the
center frequency of the spectrum-analyzer. SRCTK is used typically for bandwidths less than
300
kHz.
Bandwidths greater than 300
kHz
do not require tracking adjustment. Use SRCTK to
improve amplitude accuracy and maximize signal response. Use SRCTKPK to adjust tracking
automatically. See “SRCTKPK.
n
Programming Commands 5-517