RLPOS

Reference-Level Position

Selects the position of the reference level.
Syntax
RLPOS
XRLPOS
Item Description/Default Range
Number Any real or integer number.
0 to 8.
Step Increment: 1.
Related Commands: IP, MEASURE, NRL, RL.
Preset State: RLPOS 8.
Example
OUTPUT
718;"IP;"
OUTPUT
718;"MEASURE
NRM;"
OUTPUT
718;"AMBPL
ON;"
OUTPUT
718;"RLPOS
7;"
Initializes spectrum analyzer
Changes the measurement mode to normalized.
Activates trace normalization.
Rwitions
the
refwence
level at the seventh major gratic-
ule
division.
Description
The RLPOS command changes the position of the reference level during log display mode. The
top and bottom graticule lines correspond to 8 and 0, respectively. RLPOS must be used with
MEASURE NRM or MEASURE SR, and AMBPL ON or AMB ON. Arrows appear on the left and
right side of the screen graticule when the reference level position is changed.
Query Response
5-470
Programming Commands