SS Center Frequency Step Size
Example
10
CLEAR 718
20 OUTPUT
718;"IP;SNGLS;CF
300MHZ;SP
20MHZ;TS;"
30 OUTPUT
718;"MKPK
HI;MKRL;TS;MKF?;"
40
ENTER 718 USING
"K";Mk,freq
50
OUTPUT
718;"MKA?;"
60
ENTER 718 USING
"K";Mk-amp
70 OUTPUT
718;"SS
";Mk-freq;"HZ"
80 OUTPUT
718;"CF
UP;TS;MKPK
HI;MKA?;"
90 ENTER
718;Mk-amp1
100 PRINT "THE FUNDAMENTAL IS
";Mk-amp-Mk,amp1
110 PRINT
"dB
ABOVE THE SECOND HARMONIC"
120 END
Description
Clears the
HP-B
Initializes the spectrum
an&y-
activatessingle-sweep
mode,
changes
the center frequency, span, takes
sweep.
Finds the highest peak, changes
the reference level to the
markq
takes sweep, returns the frequency
of the
marker:
Puts the spectrum analyzer re-
sponse in the computer variable,
h!k-freq.
Returns
the amplitude of the
marker
Puts the spectrum analyzer re-
sponse in the computer variable,
h&amp.
Changes the step size to the marker
freq-u
Increases the centerfrequency, takes
sweep, puts the marker on the
highest peak and returns the am-
plitude of the marker
Puts the spectrum analyzer re-
sponse in the computer variable,
Mk-ampl.
Outputs the result.
The AUTO parameter removes SS as an active function but does not have an effect on its value.
Query Response
Programming Commands