UP

up

UP

UP

Increases the value of the active function by the applicable step size.
Syntax
(
UP
Related Commands: See the list of active functions listed in the description for UP
Example
OUTPUT
718;"IP;MKN;RB
1OKHZ;MKPK
NH;UP;"
Increases
the resolution bandwidth to
30
kHz
because
MKPK
NH does not change
the active function.
Description
Before executing UP, be sure that the function to be decreased is the active function. For
example, the programming example increases
the
resolution bandwidth, because marker peak
(MKPK) is not
an
active function.
The active functions are ACPBW, ACPSP, AT, CF, CRTHPOS, CRTVPOS, DL,
DOTDENS,
FA,
FB, FMGAIN, GD, GL, LG, MKA, MKD, MKFCR, MKN, MKPAUSE, MKPX, ML, NDB, NRL, RB,
RCLS, ROFFSET, RL, RLPOS, SAVES, SAVRCLN,
SETDATE,
SETTIME,
SP, SQLCH, SRCALC,
SRCAT, SRCPOFS, SRCPSWP, SRCPWR, SRCTK, SS, ST, TH, TVLINE, VB, VBR, and user-defined
active function specified by the ACTDEF command.
Programming Commands
5.579