Chapter 3 105

PSA Functional Tests

Amplitude Accuracy at 50 MHz

PSA Functional Tests

Figure 3-2. Amplitude Accuracy Test SetupTable3-1 Amplitude Accuracy Results
Nominal
Source
Amplitude
(dBm)
Preamp
(Option
1DS)
Attenuation
(dB) Span
(MHz)
Measured
Amplitude
Meas_amp
(dBm)
Power Met er
Amplitude
Power_m eter
(dBm)
Amplitude
Accuracy
Error
(dB)
Test Lim it
(dB)
4Off 102 ± 1.0
9Off 101 ± 1.0
−14 Off 10 0.5 ± 1.0
4 Off 20 0.1 ± 1.3
−14 Off 20 0.1 ± 1.3
4 Off 30 0.1 ± 1.3
14 Off 30 0.1 ± 1.3