Chapter 1 87
Instrument Messages
Instrument Messages
Instrument Messages
Status (ADC Align Failure) status bit only, no message
The alignment routine was unable to align the analog-to-digital converter
(ADC).
Status Align Now All Needed (Align Needed)
The instrument requires complete alignment. Press System, Alignments,
Align Now, All. On all Agilent Technologies ESA spectrum analyzer models
except Agilent Technologies E4401B and E4411B, you must connect the
AMPTD REF OUT to the INPUT with the appropriate cable to perform this
alignment. For Agilent TechnologiesE4401B and E4411B only: disconnect
any signals from the INPUT prior to performing this procedure. If this
message recurs, load defaults (System, Alignments, Load Defaults) and then
perform Alignment Now, All.
Status Align Now RF Needed (Align Now RF Needed)
The instrument requires RF alignment. Press System, Alignments, Align
Now, RF (EXT Cable). On all Agilent Technologies ESA spectrum analyzer
models except Agilent Technologies E4401B and E4411B, you must
connect the AMPTD REF OUT to the INPUT with the appropriate cable to
perform this alignment. For Agilent TechnologiesE4401B and E4411B
only: disconnect any signals from the INPUT prior to performing this
procedure.
Status Align RF Skipped (Align RF Skipped)
The RF alignment has been skipped because a 50 MHz signal was detected
at the INPUT; alignment will resume when the 50 MHz signal is removed.
The alignment will not work when there is too much input power at
50 MHz. The instrument may not continue to measure properly. To remove
the message, remove the 50 MHz inpu t signal, then perform an Align Now,
RF. Press System, Alignments, Align Now, RF. On all Agilent Technologies
ESA spectrum analyzer models except E4401B and E4411B, you must
connect the AMPTD REF OUT to the INPUT with the appropriate cable to
perform this alignment. For Agilent Technologies E4401B and E4411B
only: disconnect any signals from the INPUT prior to performing this
procedure.
If this message occurs and you are going to make a measurement near
50 MHz, select System, Alignments, Auto Align , and All but RF.
Status Align RF Skipped, Align signal amplitude too low
The RF alignment has been skipped because the calibration output power
is too low.