Chapter 2 99Functional TestsTest Equipment

Functional Tests

PSA ESA Cable, BNC (2 required) 120 cm (48 in.) BNC cable 10503A
Signal Source
PSA Synthesized Sweeper Frequency: 10MHz to 26.5 GHz
Harmonic level:< 40dBc
Amplitude range: 10 to 20dBm
Frequency Accuracy: 0.02%
83630B 83640B,
83650B
PSA Synthesized Sweeper Frequency:10MHz to 50GHz
Harmonic level:< 40dBc
Amplitude range: 10 to 20dBm
Frequency Accuracy: 0.02%
E4446A, E4448A only
83650B
ESA Synthesized Sweeper Frequency: 10 MHz to 26.5GHz
Harmonic level:< 40dBc
Amplitude range: 10 to 20dBm
Frequency Accuracy: 0.02%
83640A/B 836XX
series
Miscellaneous Equipment
PSA ESA Filter, 50 MHz Low Pass Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
0955-0306
PSA Power Meter Power Reference Accuracy: ±1.2%
Compatible with power sensor E4418B E4419B
ESA Power Meter Power Reference Accuracy: ±1.2%
Compatible with power sensor 438A
PSA ESA Power Sensor Frequency Range: 50 MHz to
26.5GHz
Amplitude Range: 25 to 10 dBm
8485A E4413A
ESA Power Sensor Frequency Range: 100kHz to
3.0GHz
Amplitude Range: 25 to 10 dBm
8485A E4413A
PSA Power Sensor Frequency Range: 50 MHz to
50GHz
E4446A, E4448A only
8487D
PSA ESA Power Splitter, 3.5 mm Nominal Insertion Loss: 6 dB
Tracking Between Port s: < 0.25 dB 11667B 11667A
PSA Power Splitter Frequency Range: 50 MHz to
50GHz
Tracking between ports: < 0.25 dB
E4446A, E4448A only
11667C
Analyzer Item Critical Specifications Recommended
Agilent Model
Alternate
Agilent
Model