Chapter 1 57
Instrument Messages
Instrument Messages
Instrument Messages
10149 Opt BAH DSP algorithm code file requires upgrade.
The Digital Signal Processing algorithm code file revision is not supported
by the currently loaded personality version. Refer to the user’s guide for
the personality in use for more informatio n on in s tallation/upgrade.
10150 Opt BAH DSP algorithm coef. file requires upgrade.
The Digital Signal Processor algorithm coefficient file revision is not
supported by the currently loaded personality version. Refer to the user’s
guide for the personality in use.
10151 Opt BAH DSP algorithm files failed to load, aborting measure.
The Digital Signal Processor algorithm files required to perform the
demodulation are corrupt and cannot be loaded properly. Reinstall
measurement personality.
10151 B7D DSP Code Fail. Reload Opt BAH. Contact your Service
Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are present in the IDSP: drive and cannot be loaded
properly into DSP RAM. Reinstall measurement personality.
10152 Lost trigger, aborting measurement.
The selected trigger source was present at the start of the measurement,
but timed out before the measure ment completed.
10153 DSP algorithm timeout, aborting measurement
The Digital Signal Processor demodulation algorithm timed-out for an
unknown reason. This message normally indicates a problem with the
modulated signal.
10154 Measurement not defined for Out of Band.
You have changed to an out-of-band frequency range. The band
measurement only operates in the selected band.
10155 No Fast ADC hardware installed. Meas unavailable.
The analyzer cannot use sweep times of less than 5 msec when Option B7D
or Option AYX is not installed. Therefore, the measurement will not be
executed.
10156 No Fast ADC hardware installed.
The analyzer cannot use sweep times of less than 5msec when Option
B7D or Option AYX is not installed resulting in all the radio standards not
being supported.
10156 The calibration data is invalid, and has been cleared.
10157 Tracking Generator hardware is not present. Meas unavailable.
The measurement requires a built-in tracking generator.