44 Chapter 1
Instrument Messages
Instrument Messages
Instrument Messages
221 Invalid option, unable to uninstall package
You have attempted to remove a personality that is not cur rently installed.
Verify command was entered correctly.
222 Command not valid when no measurement is active
Indicates that the command sent from the remote interface must be issued
while a measurement is running in the analyzer.
223 Trigger Offset unavailable in swept spans
Trigger Offset is only available in Zero Span. Refer to Trig in your
Agilent Spectrum Analyzer Users Guide for a description of this function.
224 Option not licensed.
The selected option requires a license. Refer to the installation procedures
in the users guide available for this particular option.
226 Actual Trig Delay = Remote Trig Delay + Remote Trig Offset
228 Preselector centering failed - inadequate peak
You tried to center the preselector with a trace peak of insufficient level.
You may have pressed Presel Center with the marker placed below the
signal peak.
229 3 26.5 GHz preselector bypassed, cant adjust
The microwave preselector has been bypassed. Attempts to press or send
the SCPI commands for Presel Center or Presel Adjust will result in this
error.
230 Unpreselected external mixing - no preselector
While using external mixing with an unpreselected mixer, attempts to
press or send the SCPI commands for Presel Center or Presel Adjust will
result in this error.
231 Preselector unavailable while in external mixing
The internal preselectors cannot be used while in external mixing.
232 External mixing is off, cant adjust external preselector
The preselector is unavailable when external mixing is turned off.
238 Trace smoothing: VBW filt or Average Detector; cannot use both
239 HW Diag is OFF
332 Average Type incompatible for scale.
Amplitude Scale command should be sent prior to the Average Type
command.