Chapter 1 75
Instrument Messages
Instrument Messages
Instrument Messages
10778 Valid PRACH burst not found.
This error message appears when an PRACH power measurement
(Waveform/Chip Power) is selected and a valid PRACH burst is not
found.
10801 to 10900: cdma2000 Error Messages (Option B78) An error number in this
range indicates the instrument has detected an error relating to the cdma2000 personality.
10824 Signal too noisy
10826 Input power too low
The instrument only identifies a channel as active if it meets the default
threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be chan ged using the Active Set Thresh old
function in the Meas Setup menu.
10868 Can not get long code phase (RS-232)
For MS (mobile station) measurements, the long code phase information
could not be obtained from the signal at the RS-232 port (from code domain
power measurement or composite EVM measurement).
10872 Cannot correlate to input signal
Cannot correlate to the input signa l and no active channel is found. (fro m
composite EVM measurement) An active channel must meet the default
threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be changed using the active set threshold
function in the Meas Setup menu.
10901 to 10925: NADC Error Messages (Option BAE) An error number in this range
indicates the instrument has detected an error relating to the NADC personality.
10902 Sync word not found
In an EVM measurement, the sync word is not found and the
synchronization cannot be established when Sync Word is selected in the
Burst Sync menu.
10904 Valid NADC burst not found
A valid NADC burst is not found when the Device is MS.
10906 Signal too noisy
The valid EVM measurement cannot be performed, because the input
signal is too noisy.
10908 Burst Delay out of limit for EVM (2 ms)
In an EVM measurement, the Burst Delay value must be less than 2 ms.