Chapter 4 163
ESA Functional Tests
Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B, E4407B and E4408B,
Option1DN
ESA Functional Tests
Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B, E4407B and E4408B, Option 1DN
Test Limits
Test Description

A calibrated power sensor is connected to the tracking generator output to measure the

power level at 50 MHz. The power meter is set for REL mode so that future power level

readings are in dB relative to the power level at 50 M H z . The tracking generator is then

stepped to several frequencies throughout its range. The output power difference relative

to the power level at 50 MHz is measured at each frequency and recorded.

Maximum Maximum
Flatness 10 MHz 3.5 dBm +3.5 dBm
Flatness > 10 MHz 2.5 dBm 2.5 dBm
Item Critical Specifications
(for this test) Recommended
Agilent Model
Meters
Power Meter 438A or E4418A,
E4419A
RF Power Sensor Frequency Range: 100 kHz to 3.0 GHz 8482A
Adapters
Type-N (f) to BNC
(m), 75-Ohm 1250-1534
Cables
(2) BNC, 122-cm
(48-in) 10503A