10421 Cannot lock to carrier.
This error only occurs when attempting to demodulate OFFSET QPSK
signals. It indicates that the demodulation algorithm is unable to lock to
your signal. Possible causes may be:
1. There is no carrier signal.
2. Carrier signal is present, but Demod settings do n ot ma tch th e
modulation format of the OFFSET QPSK signal being measur ed .
3. OFFSET QPSK signal is too noisy to achieve carrier lock.
10425 B7D DSP Code Install Fail Opt.231. Contact your Service
Center.
The Digital Signal Processor algorithm files required to
for the 89600 VSA SW are not present in the
analyzer. Reinstall the measurement personality and
contact your service center.
10426 Opt 231 DSP algorithm code file requires upgrade.
The Digital Signal Processing algorithm code file
revision is not supported by the currently loaded
personality version. Refer to the user’s guide for the
personality in use for more information on
installation/upgrade.
10427 Opt 231 DSP algorithm coef. file requires upgrade.
The Digital Signal Processor algorithm coefficient file
revision is not supported by the currently loaded
personality version. Refer to the user’s guide for the
personality in use for more information on
installation/upgrade.
10428 B7D DSP Code Reload Fail. Opt.231. Contact your Service
Center.
The Digital Signal Processor algorithm files required for
the 89600 VSA SW are corrupt and cannot be
loaded properly. Reinstall the measurement personality
and contact your service center.