58 Chapter 1
Instrument Messages
Instrument Messages
Instrument Messages
10158 This operation requires a measurement to be active.
10159 Entire trace is below threshold level
The measurement cannot operate properly because the trace has
completely fallen below the threshold level. Change the threshold level to
view trace.
10160 Valid burst not found
Cannot find a valid burst, either because it is not fully on the display, or
the level is too low to be detected properly.
10160 Upper Custom Mask is Invalid!
The user-specified upper custom mask cannot be resolved into a limit line.
The format may be bad.
10161 No Fast ADC hardware installed.
Cannot use sweeptimes less than 5 ms, due to the lack of a fadc, therefore
the meas will be severely restricted.
10161 Lower Custom Mask is Invalid!
The user-specified lower custom mask cannot be resolved into a limit line.
The format may be bad.
10162 Resolution BW <300 kHz.
This error message is a warning that the resolution bandwidth has been
set below 300 kHz. The test results will not meet GSM specifications.
10163 Cannot find the Power vs Time Limits File.
The limit line definition file for the GSM standards has been deleted. This
message is displayed while the Measure key is grayed out. Reinstall the
GSM measurement personality.
10164 Band Measurement not defined for Out of Band.
You are attempting to monitor the band but have set the frequency outside
the band. Reset the band for the particular standard for which you are
testing or use the channel setting which does not require a frequency to be
set. (Meas Setup, Method (Channel)).
10165 Carrier Not Present.
10166 Cannot update the list of cable types.
The cable file may have been moved or deleted accidentally. Reinstall the
measurement personality. This message is applicable to the GSM (Option
BAH) and Distance to Fault (Option 226) measurement personality
options.