64 Chapter 1
Instrument Messages
Instrument Messages
Instrument Messages
10290 Parameter unavailable in demod measurements.
You have selected (by remote SCPI command) either RF Amplitude sync or
Video trigger while running one of the demod measurements.
10291 Opt B7E RF hardware required. RF Burst unavailable.
You have selected (by remote SCPI command) RF Burst Trigger with no
digital demodulation RF card present (RF Communications Hardware
(Option B7E)).
10320 Opt 106 hardware required. Preamble sync unavailable.
You have selected (by remote SCPI command) preamble sync with no
Bluetooth FM demodulation card present (Option 106).
10321 Start Marker must be at least 1% < Stop Marker.
You have attempted to input a start marker value that will result in the
difference between the start and stop markers being less than 1%.
10322 Stop Marker must be at least 1% > Start Marker.
You have attempted to input a stop marker value that will result in the
difference between the start and stop markers being less than 1%.
10323 Unable to Calculate Result using Current Setup.
You have changed the setup parameters such that the marker li nes used to
measure the power cannot be displayed therefore accurate measurements
cannot be made.
10340 10101010 pattern not detected - results may be inaccurate.
This message is displayed if the measurement cannot detect the
10101010 pattern in the payload. The measurement will continue and
carry out the calculations on the payload data supplied, but may not be
correct.
10350 Payload data pattern 10101010 not present.
This message is displayed when the Payload Data parameter is set to
Auto and the measurement has not detected either of the required
patterns in the payload.
10351 Required payload data pattern 10101010 not present.
This message is displayed after successfully measuring and holding the
11110000 pattern if the measurement is restarte d and the 10101010 data
pattern is not detected.
10352 Required payload data pattern 11110000 not present.
This message is displayed after successfully measuring and holding the
10101010 pattern if the measurement is restarte d and the 11110000 data
pattern is not detected.