156 Chapter 4
ESA Functional Tests
Scale Fidelity
ESA Functional Tests
6. Record the marker delta reading in Table 4-7. At each new synthesized sweeper power level, press Single, Peak Search, and record the marker amplitude level.Table4-7 Scale Fidelity WorksheetSynthesized
Sweeper Level Minimum (dB) Marker
Level (dB)Maximum (dB)
Reference=______ 0 (Reference) 0 (Reference)
Reference –4 dB –5.0 –3.0
Reference –16 dB –17.40 –15.60
Reference –28 dB –29.40 –26.60
Reference –40 dB –41.40 –38.60
Reference –52 dB –53.40 –50.60
Reference –64 dB –66.0 –62.0