158 Chapter 4
ESA Functional Tests
Second Harmonic Spurious Responses
ESA Functional Tests
Figure 4-7 Equipment Se tup
Procedure
1. Connect the equipment as shown in Figure 4-7.
2. Set the synthesize d sweeper controls as follows:
Frequency, 40, MHz
POWER LEVEL, –10, dBm (50 Ω Input only)
POWER LEVEL, –4.3, dBm (75 Ω Input only)
NOTE 75 Ω Input only. Connect the minimum loss adapter between the low pass
filter and 75 Ω Input.
Additional Equipment for 75 Ω Input
Pad, minimum loss 11852B
Type-N (f), to
BNC (m) 1250-1534
Item Critical Specifications
(for this test) Recommended
Agilent Model