114 Chapter 3

PSA Functional Tests

Frequency Response (Flatness)

PSA Functional Tests
Table3-5 Frequency Response (Flatness) Results
Center
Frequency
Analyzer
Amplitude
Measamp
Power Meter
Measurement
Powermeter
Meas Error
Measerror
Flatness
Normalized
to 50 MHz
FlatNorm
Flatness Error
Test Li mits
(dB)
50 MHz 0 Ref
1 GHz ±1.5
2 GHz ±1.5
3 GHz ±1.5
3.5 GHz ±2.5
5 GHz ±2.5
6 GHz ±2.5
7 GHz ±3.0
10 GHz ±3.0
13 GHz ±3.0
14 GHz ±3.0
16 GHz ±3.0
22 GHz ±3.0
23 GHz ±3.5
25 GHz ±3.5
26.5 GHz ±3.5
28 GHz ±3.5
31.5 GHz ±3.5
42 GHz ±3.5
48 GHz ±3.5