146 Chapter 4

ESA Functional Tests

Reference Level Accuracy

ESA Functional Tests
Reference Level Accuracy
Test Limits
Test Description

A 50 MHz CW signal is applied to the Input of the analyzer. The amplitude of the source

and the analyzers reference level are decreased in 10 dB steps. The analyzer marker

functions are used to measure the amplitude difference between steps. Reference Level

Accuracy is tested in both Log and Linear Scale Mod e s. Most of t he e rr o r i s co ntributed

from the output attenuator inaccuracy in the synthesized sweeper and not the analyzer.

Reference Level Minimum
(dB) Maximum
(dB)

dBm dBmV

30 21.76 Reference Reference
20 31.76 1.40 1.40
10 41.76 1.40 1.40
40 11.76 1.40 1.40
50 1.76 1.40 1.40
60 8.24 1.40 1.40
70 18.24 2.0 2.0
Item Critical Specifications
(for this test) Recommended
Agilent Model
Signal Sources
Synthesized Sweeper Output Level Accuracy
0 to 15 dBm: ±1.0 dB
16 dBm to 63 dBm: ±1.4 dB
64 dBm:2.0 dB
8340A/B or 836XX
Series
Adapters
Type-N (m), to
BNC f) 1250-1476
Cables
Type-N, 152-cm
(60-in) 11500D
BNC, 122-cm (48-in) 10503A
Additional Equipment for 75-Ohm Input
Pad, minimum loss 11852B