Chapter 1 35
Instrument Messages
Instrument Messages
Instrument Messages
occurred (see IEEE 488.2, 6.3.2.7). For example, a query was followed by
DAB or GET before a response was completely sent.
400 Query Error
This is a generic query error for devices that cannot detect more specific
errors. The code indicates only that a query error as defined in IEEE
488.2, 11.5.1.1.7 and 6.3 has occurred.
399 to 300: Device-Specific Errors
NOTE Error numbers are displayed in the error queue, not on the display.
To see an error number, press System, Show Errors.
340 Calibration failed
The instrument requires an Align All Now. Restore the alignment by
pressing System, Alignments, Align All Now.
330 Self-test failed; EEPROM checksum for <card>
The card identification header for a hardware card is incorrect. If the card
is not properly identified, the instrument is likely to be non-functional.
Report this error to the nearest Agilent Technologies sales or service office.
321 Out of memory
An internal operation needed more memory than was available. Report
this error to the nearest Agilent Technologies sales or service office.
310 System error; Pretune DAC Calibration failed
System error; Pretune DAC Cal failed midpoint verification
System error
An internal system problem was detected. Report this error to the nearest
Agilent Technologies sales or service office.
300 Device-specific error
This is a generic device-dependent error for devices that cannot detect
more specific errors. The code indicates only that a device-dependent er ror
as defined in IEEE 488.2, 11.5.1.1.6 has occurred. Report this error to the
nearest Agilent Technologies sales or service office.
299 to 200: Execution Errors
NOTE Error numbers are displayed in the error queue, not on the display.
To see an error number, press System, Show Errors.