occurred (see IEEE 488.2, 6.3.2.7). For example, a query was followed by
DAB or GET before a response was completely sent.
−400 Query Error
This is a generic query error for devices that cannot detect more specific
errors. The code indicates only that a query error as defined in IEEE
488.2, 11.5.1.1.7 and 6.3 has occurred.
−399 to −300: Device-Specific ErrorsNOTE Error numbers are displayed in the error queue, not on the display.
To see an error number, press System, Show Errors.
−340 Calibration failed
The instrument requires an Align All Now. Restore the alignment by
pressing System, Alignments, Align All Now.
−330 Self-test failed; EEPROM checksum for <card>
The card identification header for a hardware card is incorrect. If the card
is not properly identified, the instrument is likely to be non-functional.
Report this error to the nearest Agilent Technologies sales or service office.
−321 Out of memory
An internal operation needed more memory than was available. Report
this error to the nearest Agilent Technologies sales or service office.
−310 System error; Pretune DAC Calibration failed
System error; Pretune DAC Cal failed midpoint verification
System error
An internal system problem was detected. Report this error to the nearest
Agilent Technologies sales or service office.
−300 Device-specific error
This is a generic device-dependent error for devices that cannot detect
more specific errors. The code indicates only that a device-dependent er ror
as defined in IEEE 488.2, 11.5.1.1.6 has occurred. Report this error to the
nearest Agilent Technologies sales or service office.
−299 to −200: Execution ErrorsNOTE Error numbers are displayed in the error queue, not on the display.
To see an error number, press System, Show Errors.