Agilent Technologies Z5623A manual Guaranteed Performance, Parameter Specification

Models: Z5623A

1 138
Download 138 pages 46.21 Kb
Page 12
Image 12

Instrument Description

Specifications

Guaranteed Performance

Table 1-1

Agilent Z5623A Option H48 Guaranteed Performance Specs

 

 

 

 

 

 

Parameter

 

Specification

 

 

 

 

 

 

Frequency Range

300 kHz to 9.0 GHz

 

 

 

 

 

 

Crosstalka

1) –110 dBb

 

 

 

2)

–105 dBc

 

 

 

3)

–95 dBd

 

 

 

4)

–90 dBe

 

 

Return Loss (port active)f

1) 24 dBb

 

 

 

2)

18 dBc

 

 

 

3)

14 dBd

 

 

 

4)

9 dBe

 

 

Return Loss (port off)g

1) 26 dBb

 

 

 

2)

20 dBc

 

 

 

3)

16 dBd

 

 

 

4)

14 dBe

 

 

Insertion Lossh

1) 1.5 dBb

 

 

 

2)

2.0 dBc

 

 

 

3)

2.5 dBd

 

 

 

4)

3.5 dBe

 

 

Switch Lifetime

5 million cycles

 

 

 

 

 

 

Maximum Input Power

1 Watt (RF + DC)i

 

a.Between any two non-connected signal paths b. Band 1 (300 kHz to 1.3 GHz)

c. Band 2 (1.3 GHz to 3.0 GHz) d. Band 3 (3.0 GHz to 6.0 GHz) e. Band 4 (6.0 GHz to 9.0 GHz)

f.When using an external termination on the Reflection or Transmission port

g.When using one of the 50 Ω internal terminations

h.From any test set port to the Reflection or Transmission port i. Set by the power handling capability of the 50 Ω internal

terminations