Agilent Technologies Z5623A manual Completed Limit Table for Return Loss Port Off

Models: Z5623A

1 138
Download 138 pages 46.21 Kb
Page 93
Image 93

Performance Verification

Setting Up Limit Testing

Figure 5-24 Completed Limit Table for Return Loss (Port Off)

Step 4. Click Trace on the menu bar. When the Trace menu appears, scroll down to the Limit Test... button and click it as shown in Figure 5-8. The Limit Test window shown in Figure 5-9will pop up.

Step 5. Click the Hide Table button, then click the OK button.

Saving the Limit Test File

Step 1. Click File on the menu bar. When the File menu appears, scroll down to the Save As... button and click it. See Figure 5-14.

Step 2. When the Save As window comes up as shown in Figure 5-15, type a file name that includes the words, “return loss,” the words, “port off”, and the date of the calibration listed in Table 4-1 on page 4-13. For example, if the calibration file is named, “cal_06_17_01”, an appropriate file name for this step would be, “return_loss_port_off_with_ 06_17_01_cal”. See Figure 5-16. After entering the file name, record it in Table 5-3 on page 5-7. Next, click the Save button.

Recalling the Limit Test File

Recall the limit test file that you have just saved. (Consult “How to

Recall a Calibration” on page 4-14.) This will put it on the “Quick

Recall” list.

Agilent Technologies Z5623A Option H48

5 -25

Page 93
Image 93
Agilent Technologies Z5623A manual Completed Limit Table for Return Loss Port Off, Recalling the Limit Test File