Agilent Technologies Z5623A manual Cal File Highlighted

Models: Z5623A

1 138
Download 138 pages 46.21 Kb
Page 61
Image 61

Calibrating the Network Analyzer

Renaming the Calibration File

Step 9. Move the cursor over the filename that was copied down in Step 16 of “Performing the Calibration” on page 4-6. (If you were not able to copy down the filename, select the “.cst” file whose date and time correspond to when you completed the calibration.) Do not click the filename you have selected. Just keep the cursor positioned over the filename for about two seconds (until it is highlighted as shown in Figure 4-18).

Figure 4-18 Cal File Highlighted

Step 10. Move the cursor off the selected filename (either to the left or to the right). The selected filename will remain highlighted.

Agilent Technologies Z5623A Option H48

4-11

Page 61
Image 61
Agilent Technologies Z5623A manual Cal File Highlighted