Agilent Technologies Z5623A manual General Troubleshooting Notes

Models: Z5623A

1 138
Download 138 pages 46.21 Kb
Page 127
Image 127

WARNING

WARNING

CAUTION

CAUTION

Service

Troubleshooting

Troubleshooting

This section contains procedures for troubleshooting the test set to assembly level only. These procedures should enable you to determine whether the power supply, front panel, or main switch board needs replacing. Refer to the system block diagram (Figure 8-1 on page 8-2) as an aid in troubleshooting.

General Troubleshooting Notes

Always turn the instrument power off before removing or installing an assembly.

Some parts in the instrument have sharp edges. Work carefully to avoid injury.

If you need to disassemble the instrument, be sure to work at an antistatic workstation and use a grounded wrist strap. This will protect the instrument from electrostatic discharges (ESD) that could damage the instrument.

After identifying a defective assembly or board, inspect it for obvious, easy-to-fix defects such as bent pins on ICs or cold solder joints. Repair of these defects by qualified personnel will restore the instrument to normal operation much more rapidly than ordering a replacement assembly or board.

Troubleshooting Power Supply Problems

Turn the instrument on. Check the condition of the LCD on the front panel:

If it is off, there is still a possibility that the power supply is not supplying the necessary +24V, +12V, and +5V to the main board.

If the LCD is off, check the main fuse located in the power supply filter at the rear of the instrument.

If the LCD is still off, check the cable between the main board and front panel board.

Finally, disconnect the DC power cable from the power supply to the main switch board and measure the voltages. They should be +15V, +12V, and +5V. If not, replace the power supply.

Agilent Technologies Z5623A Option H48

8 -5

Page 127
Image 127
Agilent Technologies Z5623A manual General Troubleshooting Notes, Troubleshooting Power Supply Problems