WARNING
WARNING
CAUTION
CAUTION
Service
Troubleshooting
Troubleshooting
This section contains procedures for troubleshooting the test set to assembly level only. These procedures should enable you to determine whether the power supply, front panel, or main switch board needs replacing. Refer to the system block diagram (Figure
General Troubleshooting Notes
Always turn the instrument power off before removing or installing an assembly.
Some parts in the instrument have sharp edges. Work carefully to avoid injury.
If you need to disassemble the instrument, be sure to work at an antistatic workstation and use a grounded wrist strap. This will protect the instrument from electrostatic discharges (ESD) that could damage the instrument.
After identifying a defective assembly or board, inspect it for obvious,
Troubleshooting Power Supply Problems
Turn the instrument on. Check the condition of the LCD on the front panel:
•If it is off, there is still a possibility that the power supply is not supplying the necessary +24V, +12V, and +5V to the main board.
•If the LCD is off, check the main fuse located in the power supply filter at the rear of the instrument.
•If the LCD is still off, check the cable between the main board and front panel board.
•Finally, disconnect the DC power cable from the power supply to the main switch board and measure the voltages. They should be +15V, +12V, and +5V. If not, replace the power supply.
Agilent Technologies Z5623A Option H48 | 8 |