Calibrating the Network Analyzer

Performing the Calibration

Step 2. Press the Preset button on the Network Analyzer.

Step 3. Verify that the Ready light on the ECal Module is green.

Step 4. On the network analyzer, click Calibration on the menu bar, then click the Calibration Wizard button. See Figure 4-2.

Figure 4-2 Calibration Menu on Network Analyzer

Step 5. When the window shown in Figure 4-3comes up, click the Cal Type button. The Cal Type window shown in Figure 4-4will then pop up.

Figure 4-3 Cal Wizard Window

The currently selected calibration type is “None”.

Agilent Technologies Z5623A Option H48

4-3

Page 53
Image 53
Agilent Technologies Z5623A manual Calibration Menu on Network Analyzer